Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Year range
1.
Rev. odontol. UNESP (Online) ; 44(2): 67-73, Mar-Apr/2015. tab, ilus
Article in Portuguese | LILACS | ID: lil-746236

ABSTRACT

Objetivo: Analisar a resistência à fratura por fadiga dos conjuntos implante-pilar, nos tipos reto e angulado, submetidos a cargas cíclicas similares às mastigatórias. Material e método: Foram utilizados 32 implantes (3,75 × 11 mm) com sistema cone Morse e 32 pilares, separados em dois grupos: pilares retos e angulados (n=16) (Neodent, Curitiba, PR, Brasil). Os conjuntos foram submetidos a testes cíclicos em equipamento servo-hidráulico, fixando-se o número de ciclos em cinco milhões. Foram avaliados: número de ciclos, carga e momento de força das amostras. O Teste de Fisher e o Teste ANOVA foram aplicados (p<0,005). Amostras fraturadas foram analisadas em microscopia eletrônica de varredura (MEV). Resultado: Dos pilares retos, quatro resistiram ao número de ciclos estabelecidos, suportando cargas entre 470 N e 510 N. No grupo dos angulados, nove amostras resistiram a cargas entre 570 N e 890 N. Quanto às amostras fraturadas abaixo do número de ciclos, no grupo pilar reto, nove fraturaram com cargas entre 470 N e 630 N. No grupo pilar angulado, cinco amostras fraturaram com cargas entre 760 N e 890 N. Foi calculado, para cada conjunto implante-pilar, o momento de força (Médias: Grupo reto - 4.335,2 Nmm, Grupo angulado - 3.923,3 Nmm). Conclusão: Os dois tipos de pilares (retos e angulados), em condições in vitro, comportaram-se estatisticamente de forma semelhante, sendo aceita a hipótese nula de que não há diferença de resistência à fratura entre os grupos.


Objective: The aim of this study was to verify the fatigue fracture resistance of the implant-abutment joint, in straight and angled types, subject to cyclic loads similar to chewing. Material and method: It was used 32 implants system (3.75 × 11mm) with Morse taper and 32 abutment, divided in two groups: abutment straight and angled (n = 16) (Neodent, Curitiba, PR, Brazil). The sets were submitted to cyclic testing using servo-hydraulic equipment, fixing the number of cycles at 5 million. It was evaluated the number of cycles, load and moment force of the samples. The Fisher test and ANOVA were applied (p<0,005). Fractured samples were analyzed in the scanning electron microscopy (SEM). Result: Straight abutment, four resisted the cycles number established, supporting loads between 470N and 510N. Angled abutment: nine resisted to loads between 570N and 890N. Regarding the fractured samples below the number of cycles, in the straight abutment, nine fractured with loads between 470N and 630N. In the angled abutment, five fractured with loads ranging from 760N and 890N. The moment of the force was calculated for each joint implant-abutment (Averages: Straight Group - 4335,23Nmm, angled Group - 3923,37Nmm). Conclusion: Two types of abutments (straight and angled), in vitro conditions, statistically behaved similarly, and accepts the null hypothesis that there is no difference in fracture resistance between groups. .


Subject(s)
Microscopy, Electron, Scanning , Analysis of Variance , Dental Implantation, Endosseous , Mechanical Phenomena , Flexural Strength , Titanium , In Vitro Techniques , Mouth Rehabilitation
SELECTION OF CITATIONS
SEARCH DETAIL